BIOLOGICAL SAFETY CABINET A2 – Model BSC-IIA2-PRIVILEGE

Description
Protection for the user, environment, and sample when working with biological agents within levels 1, 2, and 3. Biological safety cabinet type: CLASS II according to EN 12469. Complies with the main requirements of NSF/ANSI 49, JIS K3800, SFDA YY-0569

Description

AIR:

  • Partial air recirculation in accordance with regulations: 30% exhaust and 70% recirculation.
  • The work area maintains laminar flow conditions with ISO 4 air purity (according to UNE EN ISO 14644) or Class 10 (according to Fed. Std. 209).
  • The recirculated air is filtered through a ULPA U15 absolute filter (99.9995% MPPS efficiency) or HEPA H14 filter (99.995% MPPS efficiency), in accordance with UNE EN 1822, and is delivered to the work area at a velocity between 0.25 and 0.50 m/s (in accordance with UNE EN ISO 14644).
  • The air expelled to the outside is filtered by a dedicated ULPA U15 or HEPA H14 absolute filter in accordance with the UNE EN 1822 standard.
  • Protective barrier in the work area with a velocity greater than 0.50 m/s.

 

CONSTRUCTION:

  • Work area constructed of AISI 304 stainless steel. Optional: AISI 316L.
  • The cabinet has a depth of 795 mm, facilitating passage through doors to its installation location.
  • Sliding front glass panel with an alarm position. Raises or lowers manually or electronically. Features an anti-pinch function with obstacle detection during glass closure.
  • The safety opening glass is 5 mm thick with UV protection and anti-reflective coating, with an incline of up to 10° to facilitate viewing of the samples and provide a full view of the work surface.
  • Glass with a special guide that improves sealing and safety during sliding.
  • Microprocessor-controlled cabinet with upgradeable FLASH memory and external display, easy viewing of cabinet control parameters (speeds, flow rates, plenum pressure, exhaust barriers, humidity, temperature, etc.), touch panel, easy operation, security PASSWORD.
  • Alarms for detecting problems in the air supply system and the negative pressure protection barrier, as well as door openings, via voice alerts and/or visual indicators on the screen.
  • Pre-installation for decontamination with Vaporized Hydrogen Peroxide (VHP)
  • Adjustable white LED lighting and UV-free.
  • Operating voltage 110-220 V, 50/60 Hz. Maximum current consumption 3 A (approx. 450 W).
  • The interior of the cabinet includes a UV light; 2 IP 55-rated 110-220 volt electrical outlets for 1000 W each (total 2,000 W).

 

 

“All information, specifications, dimensions, and electrical data shown in TDI’s commercial documents and data sheets are for guidance only and are not binding; therefore, they are subject to change according to each customer’s needs and provided that applicable regulations permit such changes.”

MODEL BSC-IIA2-1100P BSC-IIA2-1300P BSC-IIA2-1400P BSC-IIA2-1500P BSC-IIA2-1800P
Internal dimensions mm (WxDxH) 950*645*600 1200*645*600 1300*645*600 1400*645*600 1700*645*600
External cabin dimensions mm (WxDxH) 1050*795*1470 1300*795*1470 1400*795*1470 1500*795*1470 1800*795*1470
External dimensions of standard support table – includes: 90 mm in height: casters 1050*730*714 1300*730*714 1400*730*714 1500*730*714 1800*730*714
Tested opening Safety height 200 mm (8″)
Maximum opening 440 mm
Work area height (includes 90 mm casters) 892 mm
Control system Monitor 7″ Color LCD Touchscreen
Alarm Abnormal airflow speed, ULPA/HEPA filter failure, UV lamp failure, front window at an unsafe height.
Airflow System Airflow Mode 70% air recirculation, 30% outdoor air intake
Downward Air Velocity 0.35 m/s ± 0.025 m/s
Barrier Speed > 0.52 m/s ± 0.025 m/s
Inlet flow rate 355 m³/h

(209 CFM)

450 m³/h

(265 CFM)

486 m³/h

(286 CFM)

524 m³/h

(309 CFM)

636 m³/h

(375 CFM)

Downward Flow 718 m³/h

(423 CFM)

907 m³/h

(534 CFM)

983 m³/h

(579 CFM)

1,058 m³/h (623 CFM) 1,285 m³/h (757 CFM)
Exhaust Flow Rate 355 m³/h

(209 CFM)

450 m³/h

(265 CFM)

486 m³/h

(309 CFM)

524 m³/h

(309 CFM)

636 m³/h

(375 CFM)

Filtration System ULPA or HEPA Filters

(Exhaust and Supply)

UNE EN 1822 Filters  2-piece, ≥99.9995%, ≥99.995% efficiency at 0.1 ~ 0.3μm, with filter life indicator
Air Cleanliness Classification Class 100 (ISO 14644-1 Class 4)
Cabinet Construction Main Body Electro-galvanized steel with antimicrobial powder coating
Work Area 304 stainless steel (optional AISI 316L)
Front Glass Motorized, double-layered laminated tempered glass ≥ 5mm. UV-resistant.
Electrical system UV lamp 20W x 1 30W x 1 30Wx1 30Wx1 40Wx1
254-nanometer emission for more efficient decontamination, service life indicator, and timer.
LED lamp 16Wx2 20Wx2 20Wx2 20Wx2 20Wx2
Maximum Power Consumption 750W (Does not include power consumption from the power cord)
Operating Power Consumption 450W
Fan Power 450W
Voltage AC 220V ±10%, 50/60 Hz; 110 V ±10%, 60 Hz
Lighting ≥1000 lx
Noise Level ≤55 dB(A)
Vibration Performance: Amplitude ≤ 5 µm (rms)
Standard Accessories 1. Dimmable LED lights;  2. Air velocity sensor;  3. ULPA filters;  4. LCD touchscreen; 5. LED lamp x 2;  6. UV lamp x 1; 7. IP55-rated waterproof outlets x 2;  8. Motorized front window; 9. Solenoid valve control; 10. RS485 interface; 11. USB port; 12. Foot pedal for front window.
Optional Accessories 1. LED lighting, UGR rating <16; 2. Gas valve; 3. Vacuum valve; 4. Bar with hooks; 5. HEPA filters; 6. Base stand; 7. Electronically height-adjustable base stand; 8. Carbon filter
Net Weight (kg) 224 245 263 290 370
Gross Weight (kg) 267 290 313 380 500
Shipping dimensions mm (WxDxH) 1250*1080*1720 1450*1100*1720 1550*1120*1720 1650 x 1120 x 1720 2050*1220*1870
Shipping volume (m3 ) 2.3 2.7 3 3.2 4.7
Description Part Number
Plenum for dual ULPA or HEPA filtration in the exhaust system, providing extra safety when the unit expels air into the laboratory. Request a quote
Third filtration stage, with an additional activated carbon filter for aerosols / removal of volatile chemical compounds and odors.

Different types of activated carbon.

Request price
Side windows. Request price
High-End cabinet model option: Hinged front glass for full opening of the work area, XL height of 750 mm. Request price
Standard support table with casters and brakes. 440206-BASE-BSC-XXXX
Electric height-adjustable support table. 440206-MESAREGALTXXX
Bell-type connection for safe extraction. 00-CAMPA-EXT SEG
Check valve. 79-CAJANTR
1” connector for decontamination with vaporized hydrogen peroxide (VHP) in 316L stainless steel 00-VHP-CSB
Solenoid valve for gas tap, controlled from the cabin panel. 32-GE08150
Taps for gas, vacuum, air… 77-006831, 77-006841
LED light with a UGR rating of < 16 (UGR – Unified Glare Rating) Request a quote
CODE DESCRIPTION CABINS
440206-BSC-IIA2-1100P PRIVILEGE. C10 HIGH MICROBIOLOGICAL SAFETY CABINET MOD. BSC-IIA2-1100 EN 12469
440206-BSC-IIA2-1300P PRIVILEGE. HIGH MICROBIOLOGICAL SAFETY CABINET C10 MOD. BSC-IIA2-1300 EN 12469
440206-BSC-IIA2-1400P PRIVILEGE. C10 HIGH MICROBIOLOGICAL SAFETY CABINET MOD. BSC-IIA2-1400 EN 12469
440206-BSC-IIA2-1500P PRIVILEGE. C10 HIGH MICROBIOLOGICAL SAFETY CABINET MOD. BSC-IIA2-1500 EN 12469
440206-BSC-IIA2-1800P PRIVILEGE. C10 HIGH MICROBIOLOGICAL SAFETY CABINET MOD. BSC-IIA2-1800 EN 12469

 

CODE DESCRIPTION ACCESSORIES AND OPTIONS
440206-BSC-IIA2-XXXX-HE FRONT GLASS WITH HINGES, FULL OPENING WORK AREA, XL HEIGHT OF 750MM
440206-BASE-BSC-XXXX STANDARD TABLE FOR BSC MODEL CABINS
440206-MESAREGALTXXX ELECTRIC HEIGHT-ADJUSTABLE TABLE FOR CABINS – check sizes
00-CAMPA-EXT SEG Fume Hood for Safe Extraction in BSC Biological Safety Cabinets
79-CAJANTR 4mm THICK METHACRYLATE NON-RETURN BOX, 330mm X 150mm
32-GE08150 MADA GAS SOLENOID VALVE, AUTOMATIC RESET, N.C., 230VAC, 1/2″, Pmax 6 bar, RP
00-VHP-CSB 1” CONNECTOR FOR DECONTAMINATION WITH VAPORIZED HYDROGEN PEROXIDE – VHP 316L STAINLESS STEEL
77-006831 WALL-MOUNTED STRAIGHT GAS TAP (LIGHT GRAY WITH YELLOW KNOB, FOR NATURAL GAS, 125MM THREAD)
77-006841 WALL-MOUNTED STRAIGHT VACUUM VALVE (LIGHT GREY, FOR VACUUM UP TO 1 mbar, WITH SPECIAL 125MM THREAD)

Ask for information

Ask for more information about:
BIOLOGICAL SAFETY CABINET A2 – Model BSC-IIA2-PRIVILEGE

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.
Description
Protection for the user, environment, and sample when working with biological agents within levels 1, 2, and 3. Biological safety cabinet type: CLASS II according to EN 12469. Complies with the main requirements of NSF/ANSI 49, JIS K3800, SFDA YY-0569

Ask for information

Ask for more information about:
BIOLOGICAL SAFETY CABINET A2 – Model BSC-IIA2-PRIVILEGE

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.