Description
AIR:
- Partial air recirculation in accordance with regulations: 30% exhaust and 70% recirculation.
- The work area maintains laminar flow conditions with ISO 4 air purity (according to UNE EN ISO 14644) or Class 10 (according to Fed. Std. 209).
- The recirculated air is filtered through a ULPA U15 absolute filter (99.9995% MPPS efficiency) or HEPA H14 filter (99.995% MPPS efficiency), in accordance with UNE EN 1822, and is delivered to the work area at a velocity between 0.25 and 0.50 m/s (in accordance with UNE EN ISO 14644).
- The air expelled to the outside is filtered by a dedicated ULPA U15 or HEPA H14 absolute filter in accordance with the UNE EN 1822 standard.
- Protective barrier in the work area with a velocity greater than 0.50 m/s.
CONSTRUCTION:
- Work area constructed of AISI 304 stainless steel. Optional: AISI 316L.
- Sliding front glass panel with alarm position. Raises or lowers manually or electronically. Features an anti-pinch function with obstacle detection during glass closure.
- The glass features a special guide that improves sealing and safety during sliding.
- Microprocessor-controlled cabinet with updatable FLASH memory and external display for easy viewing of the cabinet’s control parameters (speeds, flow rates, plenum pressure, ejection barriers, humidity, temperature, etc.).
- The cabinet is equipped with an easy-to-use touchscreen panel with a security PASSWORD for controlling and powering on the unit.
- Alarms for detecting problems in the air supply system and the negative pressure protection barrier, as well as door opening, via voice alerts and visual indicators on the screen.
- Pre-installation for decontamination with Vaporized Hydrogen Peroxide (VHP)
- Adjustable white LED lighting and UV-free.
- Operating voltage 110-220 V, 50/60 Hz. Maximum current consumption 3 A (approx. 750 W).
- The interior of the cabinet includes a UV light; 2 IP 55-rated 110-220 volt electrical outlets for 1,000 W each (total 2,000 W).
“All information, specifications, dimensions, and electrical data shown in TDI’s commercial documents and data sheets are for guidance only and are not binding; therefore, they are subject to change according to each customer’s needs and provided that applicable regulations permit such changes.”