Cabinets and Air

Treatment Equipment

BSC-IIA2-700P

BIOSAFETY CABINET A2 – Model BSC-PRIVILEGE-700A2

Description
Protection for the user, environment, and sample when working with biological agents within levels 1, 2, and 3. Biological safety cabinet type: CLASS II according to EN 12469. Complies with the main requirements of NSF/ANSI 49, JIS K3800, SFDA YY-0569

Description

AIR:

  • Partial air recirculation in accordance with regulations: 30% exhaust and 70% recirculation.
  • The work area maintains laminar flow conditions with ISO 4 air purity (according to UNE EN ISO 14644) or Class 10 (according to Fed. Std. 209).
  • The recirculated air is filtered through a ULPA U15 absolute filter (99.9995% MPPS efficiency) or HEPA H14 filter (99.995% MPPS efficiency), in accordance with UNE EN 1822, and is delivered to the work area at a velocity between 0.25 and 0.50 m/s (in accordance with UNE EN ISO 14644).
  • The air expelled to the outside is filtered by a dedicated ULPA U15 or HEPA H14 absolute filter in accordance with the UNE EN 1822 standard.
  • Protective barrier in the work area with a velocity greater than 0.50 m/s.

 

CONSTRUCTION:

  • Work area constructed of AISI 304 stainless steel. Optional: AISI 316L.
  • Sliding front glass panel with alarm position. Raises or lowers manually or electronically. Features an anti-pinch function with obstacle detection during glass closure.
  • The glass features a special guide that improves sealing and safety during sliding.
  • Microprocessor-controlled cabinet with updatable FLASH memory and external display for easy viewing of the cabinet’s control parameters (speeds, flow rates, plenum pressure, ejection barriers, humidity, temperature, etc.).
  • The cabinet is equipped with an easy-to-use touchscreen panel with a security PASSWORD for controlling and powering on the unit.
  • Alarms for detecting problems in the air supply system and the negative pressure protection barrier, as well as door opening, via voice alerts and visual indicators on the screen.
  • Pre-installation for decontamination with Vaporized Hydrogen Peroxide (VHP)
  • Adjustable white LED lighting and UV-free.
  • Operating voltage 110-220 V, 50/60 Hz. Maximum current consumption 3 A (approx. 750 W).
  • The interior of the cabinet includes a UV light; 2 IP 55-rated 110-220 volt electrical outlets for 1,000 W each (total 2,000 W).

 

“All information, specifications, dimensions, and electrical data shown in TDI’s commercial documents and data sheets are for guidance only and are not binding; therefore, they are subject to change according to each customer’s needs and provided that applicable regulations permit such changes.”

MODEL BSC-IIA2-700P
External dimensions (W x D x H) mm 710 x 760 x 1430 WITHOUT support table

710 x 760 x 2050 WITH support table

Internal dimensions (W x D x H) mm 610 x 550 x 600
Tested opening Safety height 200 mm (8″)
Maximum opening 370 mm
Work area height 750 mm
Control system Monitor 7″ Color LCD Touchscreen
Alarm Abnormal airflow speed, ULPA filter failure, UV lamp failure, front window at an unsafe height.
Airflow System Airflow Mode 70% air recirculation, 30% outdoor exhaust
Downward Air Velocity 0.35 m/s ± 0.025 m/s
Barrier velocity > 0.52 m/s ± 0.025 m/s
Air intake flow rate 240 m³/h (141 CFM)
Downflow air flow rate 560 m³/h (330 CFM)
Exhaust air flow rate 240 m³/h (141 CFM)
Filtration system ULPA or HEPA filters

(Exhaust and Supply)

UNE EN 1822 ULPA or HEPA filters, 2 pieces, ≥99.9995%, ≥99.995% efficiency at 0.1–0.3 μm, filter life indicator,
Air purity classification Class 100 (ISO 14644-1 Class 4)
Cabinet construction Main body Electro-galvanized steel with antimicrobial powder coating
Work area 304 stainless steel (Optional AISI 316)
Front glass Motorized, double-layered laminated tempered glass ≥ 5 mm. UV-resistant.
Electrical System UV lamp 15W x 1
254-nanometer emission for more efficient decontamination, UV life indicator, UV timer.
LED Lamp 10W x 2
Maximum Power Consumption 350W (Does not include power consumption from the power outlet)
Fan Power 300W
Voltage AC 220V ±10%, 50/60 Hz; 110 V ±10%, 60 Hz
Lighting ≥1000 lx
Noise Level ≤55 dB(A)
Vibration Performance: Amplitude ≤ 2 µm (rms)
Standard Accessories 1. Dimmable LED lights;  2. Air velocity sensor;  3. ULPA filters;  4. LCD touchscreen; 5. LED lamp x 2;  6. UV lamp x 1; 7. IP55-rated waterproof connectors x 2;  8. Motorized front window; 9. Solenoid valve control; 10. RS485 interface; 11. USB port; 12. Foot pedal for front window.
Optional Accessories 1. LED lighting, UGR rating <16; 2. Gas valve; 3. Vacuum valve; 4. Bar with hooks; 5. HEPA filters; 6. Base stand; 7. Carbon filter; 8. Electronically height-adjustable base stand.
Net Weight (kg) 130 (without support table),                         140 (with support table)
Gross Weight (kg) 165 (WITHOUT Support Table),                         175 (WITH Support Table)
Shipping Dimensions (W x D x H) (mm) 850 x 990 x 1640
Shipping Volume (m3 ) 1.4
Description Reference
Plenum for dual ULPA or HEPA filtration at the exhaust, for extra safety when the unit expels air into the laboratory. Request a quote
Third filtration stage, with an additional activated carbon filter for aerosols / removal of volatile chemical compounds and odors.

Different types of activated carbon.

Request price
Side windows. Request price
Front glass panel with hinges for full opening of the work area, XL height of 750 mm. High End. Request price
Standard support table with casters and brakes. 440206-BASE-BSC-XXXX
Electric height-adjustable support table. 440206-MESAREGALTXXX
Bell-type connection for safe extraction. 00-CAMPA-EXT SEG
Check valve. 79-CAJANTR
1” connector for decontamination with vaporized hydrogen peroxide (VHP) in 316L stainless steel 00-VHP-CSB
Solenoid valve for gas tap, controlled from the cab panel. 32-GE08150
Taps for gas, vacuum, air… 77-006831, 77-006841
LED light with UGR rating < 16 (UGR – Unified Glare Rating) ask for price
CODE DESCRIPTION
440206-BSC-IIA2-700P PRIVILEGE. C10 HIGH MICROBIOLOGICAL SAFETY CABINET MOD. BSC-IIA2-700 EN 12469

 

CODE DESCRIPTION OF ACCESSORIES AND OPTIONS
440206-BASE-BSC STANDARD TABLE FOR BSC MODEL CABIN
440206-MESAREGALTXXX ELECTRIC HEIGHT-ADJUSTABLE TABLE FOR CABINS
00-CAMPA-EXT SEG Fume Hood for Safe Extraction in BSC Biological Safety Cabinets
79-CAJANTR 4mm THICK METHACRYLATE NON-RETURN BOX, 330mm X 150mm
32-GE08150 MADAS GAS SOLENOID VALVE, AUTOMATIC RESET, N.C., 230VAC, 1/2″, Pmax 6 bar RP
77-006831 WALL-MOUNTED STRAIGHT GAS TAP (LIGHT GRAY WITH YELLOW KNOB, FOR NATURAL GAS, 125MM THREAD)
77-006841 WALL-MOUNTED STRAIGHT VACUUM VALVE (LIGHT GREY, FOR VACUUM UP TO 1 mbar, WITH SPECIAL 125MM THREAD)
00-VHP-CSB 1” CONNECTOR FOR DECONTAMINATION WITH VAPORIZED HYDROGEN PEROXIDE – VHP 316L STAINLESS STEEL

Ask for information

Ask for more information about:
BIOSAFETY CABINET A2 – Model BSC-PRIVILEGE-700A2

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.
Description
Protection for the user, environment, and sample when working with biological agents within levels 1, 2, and 3. Biological safety cabinet type: CLASS II according to EN 12469. Complies with the main requirements of NSF/ANSI 49, JIS K3800, SFDA YY-0569

Ask for information

Ask for more information about:
BIOSAFETY CABINET A2 – Model BSC-PRIVILEGE-700A2

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.