Description
AIR:
- Recirculated air is filtered through a ULPA U15 absolute filter (99.9995% MPPS efficiency) or a HEPA H14 filter (99.995% MPPS efficiency), in accordance with UNE EN 1822, and is delivered to the work area at a velocity between 0.25 and 0.50 m/s (in accordance with UNE EN ISO 14644).
- 100% exhaust air. Equipped with two motors, one for supply and another external motor for complete cabin exhaust.
- The air expelled to the outside is filtered by a dedicated ULPA U15 or HEPA H14 absolute filter in accordance with the UNE EN 1822 standard.
- The work area maintains laminar flow conditions with ISO 4 air purity (according to UNE EN ISO 14644) or Class 10 (according to Fed. Std. 209).
- Protective barrier in the work area with a velocity greater than 0.50 m/s.
CONSTRUCTION:
- Work area constructed of AISI 304 stainless steel. Optional: AISI 316L.
- The cabinet has a depth of 795 mm, facilitating passage through doors to its installation location.
- Sliding front glass panel with alarm position. Raises or lowers manually or electronically. Features an anti-pinch function with obstacle detection during glass closure.
- The safety opening glass is 5 mm thick with UV protection and anti-reflective coating, with an incline of up to 10° to facilitate the user’s viewing of the samples and provide a full view of the work surface.
- Glass with a special guide that improves sealing and safety during sliding.
- Microprocessor-controlled cabinet with upgradeable FLASH memory and external display for easy viewing of the cabinet’s control parameters (speeds, flow rates, plenum pressure, exhaust barriers, humidity, temperature, etc.), user-friendly touchscreen with a security PASSWORD for controlling and powering on the unit.
- Alarms for detecting problems in the drive system and the negative pressure protective barrier, as well as door opening, via voice alerts and/or visual indicators on the screen.
- Pre-installation for decontamination with Vaporized Hydrogen Peroxide (VHP)
- Adjustable white LED lighting and UV-free.
- Operating voltage 110-220 V, 50/60 Hz. Maximum current consumption 3 A (approx. 450 W).
- The interior of the cabin includes a UV light; 2 IP 55-rated 110-220-volt electrical outlets for 1,000 W each (total 2,000 W).
- Installation requires 1 exhaust fan and its ductwork.
Optionally, the booth can be computerized for hardware integration:
- 1 USB port
- 1 HDMI port
- 1 DP Internet port
- 1 21.5-inch monitor
- Keyboard stand
- Computer stand