Cabinets and Air

Treatment Equipment

CLASS II A2 BIOSAFETY CABINET – HIGH END Model

Description
Class II biological safety cabinet in accordance with EN 12469. Meets the main requirements of NSF/ANSI 49, JIS K3800 and SFDA YY-0569. Protection of the user, the environment and the sample when working with biological agents at biosafety levels 1, 2 and 3

Description

AIR:

  • Partial air recirculation in accordance with applicable standards: 30% exhaust and 70% recirculation.
  • The work area maintains laminar airflow conditions with ISO 4 air cleanliness (per EN ISO 14644) or Class 10 (per Fed. Std. 209).
  • Recirculated air is filtered through a ULPA U15 absolute filter (99.9995% efficiency at MPPS) or a HEPA H14 filter (99.995% efficiency at MPPS), in accordance with EN 1822, and is delivered to the work area at a velocity between 0.25 and 0.50 m/s (per EN ISO 14644).
  • Air exhausted to the outside is filtered through a separate, dedicated ULPA U15 or HEPA H14 absolute filter in accordance with EN 1822.
  • Protective air barrier at the work opening with a velocity above 0.50 m/s.

 

CONSTRUCTION:

  • Work area made of AISI 304 stainless steel. Optional: AISI 316L.
  • The cabinet has a depth of 795 mm, making it easy to pass through doorways to its installation site.
  • Sliding front sash with alarm position. Raised and lowered manually or electronically. Anti-pinch function with obstacle detection when closing the sash.
  • 6 mm safety glass sash with UV protection and anti-reflective coating, tilted up to 10° for better sample visibility and a full view of the entire work surface.
  • Glass sash with special guide for improved sealing and safer sliding.
  • Allows full opening of the work area with an extra height of 750 mm, making cleaning easier and allowing access for large equipment. ** See dimensions of the extra-height model.
  • Alarms for detecting faults in the supply air system, the negative-pressure protective barrier and door opening, with voice and/or visual on-screen alerts
  • Microprocessor-controlled cabinet with updatable FLASH memory, touch panel, and easy-to-read control parameters (velocities, airflow rates, plenum pressure, exhaust barriers, humidity, temperature, etc.). Security PASSWORD.
  • Pre-installation for decontamination with Vaporized Hydrogen Peroxide (VHP)
  • Adjustable white LED lighting, UV-free.
  • Operating voltage 110-220 V, 50/60 Hz. Maximum current consumption 3 A (approx. 450 W).
  • The cabinet interior includes a UV light and 2 IP 55 electrical sockets (110-220 V), each rated for 1,000 W (2,000 W in total).

 

 

“All information, features, dimensions and electrical data shown in TDI’s commercial documents and data sheets are indicative and non-contractual and are therefore subject to change according to each customer’s needs, provided the applicable regulations so permit.”

MODEL BSC-IIA2-1100HE BSC-IIA2-1300HE BSC-IIA2-1400HE BSC-IIA2-1500HE BSC-IIA2-1800HE
Internal size mm (WxDxH) 950*645*600 1200*645*600 1300*645*600 1400*645*600 1700*645*600
Cabinet body external size mm (WxDxH) 1050*795*1470 1300*795*1470 1400*795*1470 1500*795*1470 1800*795*1470
External size of standard support stand

includes: 90 mm in height: wheels

1050*730*714 1300*730*714 1400*730*714 1500*730*714 1800*730*714
Tested opening Safety height 220 mm (8″)
Maximum opening 450 mm
Work zone height

includes 90 mm wheels

892 mm
Control system Monitor 7″ color LCD touchscreen
Alarm Abnormal airflow velocity, ULPA/HEPA filter failure, UV lamp failure, front sash at unsafe height.
Airflow System Airflow mode 70% air recirculation, 30% exhaust to outside
Downflow velocity 0.35 m/s ± 0.025 m/s
Barrier velocity > 0.52 m/s ± 0.025 m/s
Inflow volume 355m³/h

(209CFM)

450m³/h

(265CFM)

486m³/h

(286CFM)

524m³/h

(309CFM)

636m³/h

(375CFM)

Downflow volume 718m³/h

(423CFM)

907m³/h

(534CFM)

983m³/h

(579CFM)

1058m³/h (623CFM) 1285m³/h (757CFM)
Exhaust volume 355m³/h

(209CFM)

450m³/h

(265CFM)

486m³/h

(309CFM)

524m³/h

(309CFM)

636m³/h

(375CFM)

Filtration System ULPA or HEPA filters

(Exhaust and Supply)

EN 1822 filters, 2 pcs, ≥99.9995% and ≥99.995% efficiency at 0.1–0.3 μm, with filter life indicator
Air cleanliness classification Class 10 (ISO 14644-1 Class 4)
Cabinet Construction Main body Electro-galvanized steel with antimicrobial powder coating
Work area 304 stainless steel (optional AISI 316L)
Front sash Motorized, double-layer laminated tempered glass ≥ 6 mm. Anti-UV.
Electrical System UV lamp 20Wx1 30Wx1 30Wx1 30Wx1 40Wx1
254 nm emission for more efficient decontamination; lifetime indicator and timer.
LED lamp 16Wx2 20Wx2 20Wx2 20Wx2 20Wx2
Maximum consumption 750 W (excludes power consumption of socket loads)
Operating consumption 450W
Fan power 450W
Voltage AC 220 V ±10%, 50/60 Hz; 110 V ±10%, 60 Hz
Illumination ≥1000 lx
Sound emission ≤55 dB(A)
Vibration Performance: amplitude ≤ 5 µm (rms)
Standard Accessories 1. Dimmable LED lights;  2. Air velocity sensor;  3. ULPA filters;  4. LCD touchscreen; 5. LED lamp x 2;  6. UV lamp x 1; 7. IP55 sealed sockets x 2;  8. Motorized front sash; 9. Solenoid valve control; 10. RS485 interface; 11. USB port; 12. Front sash pedal.
Optional Accessories 1. UGR<16 LED lighting; 2. Gas tap; 3. Vacuum tap; 4. Hook bar; 5. HEPA filters; 6. Base stand; 7. Electrically height-adjustable base stand; 8. Carbon filter.
Net weight (kg) 224 245 263 290 370
Gross weight (kg) 267 290 313 380 500
Shipping dimensions mm (WxDxH) 1250*1080*1720 1450*1100*1720 1550*1120*1720 1650*1120*1720 2050*1220*1870
Shipping volume (m³) 2,3 2,7 3 3.2 4,7
Description Reference
Plenum for double ULPA or HEPA filtration on the exhaust, for extra safety when the unit exhausts air into the laboratory. price on request
Third filtration stage with additional activated carbon filter for aerosols / removal of volatile chemical compounds and odors.

Different types of activated carbon.

price on request
Side glass panels. price on request
Standard support stand with wheels and brakes. 440206-BASE-BSC-XXXX
Electric height-adjustable support stand. 440206-MESAREGALTXXX
Canopy (thimble) connection for safe exhaust. 00-CAMPA-EXT SEG
Check valve. 79-CAJANTR
1” connector for decontamination with vaporized hydrogen peroxide (VHP), in 316L stainless steel 00-VHP-CSB
Solenoid valve for gas tap, controlled from the cabinet panel. 32-GE08150
Taps for gas, vacuum, air, etc. 77-006831, 77-006841
UGR < 16 LED light (UGR – Unified Glare Rating) price on request

Dimensions of the HIGH END extra-height model: BSC-IIA2-HIGH END-75:

MODEL BSC-IIA2-1100HE-75 BSC-IIA2-1300HE-75 BSC-IIA2-1400HE-75 BSC-IIA2-1500HE-75 BSC-IIA2-1800HE-75
Internal size mm (WxDxH) 950*645*750 1200*645*750 1300*645*750 1400*645*750 1700*645*750
Cabinet body external size mm (WxDxH) 1050*795*1540 1300*795*1540 1400*795*1540 1500*795*1540 1800*795*1540
External size of standard support stand, includes: 90 mm in height: wheels 1050*730*714 1300*730*714 1400*730*714 1500*730*714 1800*730*714
Tested opening Safety height 220 mm (8″)
Maximum opening 450 mm
Work zone height

includes 90 mm wheels

892 mm
Net weight (kg) 234 257 273 304 390
Gross weight (kg) 277 302 326 394 520
Shipping dimensions mm (WxDxH) 1250*1080*1800 1450*1100*1800 1550*1120*1800 1650*1120*1800 2050*1220*1900
Shipping volume (m³) 2,4 2,8 3,1 3,3 4,8
CODE CABINETS DESCRIPTION
440206-BSC-IIA2-1100HE HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1100 EN 12469
440206-BSC-IIA2-1300HE HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1300 EN 12469
440206-BSC-IIA2-1400HE HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1400 EN 12469
440206-BSC-IIA2-1500HE HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1500 EN 12469
440206-BSC-IIA2-1800HE HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1800 EN 12469

 

440206-BSC-IIA2-1100HE-75 HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1100 EN 12469 – interior height 75 cm
440206-BSC-IIA2-1300HE-75 HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1300 EN 12469 – interior height 75 cm
440206-BSC-IIA2-1400HE-75 HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1400 EN 12469 – interior height 75 cm
440206-BSC-IIA2-1500HE-75 HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1500 EN 12469 – interior height 75 cm
440206-BSC-IIA2-1800HE-75 HIGH END. HIGH-SAFETY MICROBIOLOGICAL CABINET C10 MOD. BSC-IIA2-1800 EN 12469 – interior height 75 cm

 

CODE ACCESSORIES AND OPTIONS DESCRIPTION
440206-BASE-BSC-XXXX STANDARD STAND FOR BSC MODEL CABINETS
440206-MESAREGALTXXX ELECTRIC HEIGHT-ADJUSTABLE STAND FOR CABINETS – please enquire about sizes
00-CAMPA-EXT SEG CANOPY (THIMBLE) CONNECTION FOR SAFE EXHAUST OF BSC BIOLOGICAL SAFETY CABINETS
79-CAJANTR PMMA CHECK-VALVE BOX, 4 mm THICK, 330 mm X 150 mm
32-GE08150 MADAS GAS SOLENOID VALVE, AUTOMATIC RESET, N.C., 230 VAC, 1/2″, Pmax 6 bar, RP
00-VHP-CSB 1” CONNECTOR FOR DECONTAMINATION WITH VAPORIZED HYDROGEN PEROXIDE – VHP, 316L STAINLESS STEEL
77-006831 WALL-MOUNTED GAS TAP, STRAIGHT OUTLET

(LIGHT GREY, YELLOW KNOB, FOR NATURAL GAS, SPECIAL 125 MM THREAD)

77-006841 WALL-MOUNTED VACUUM TAP, STRAIGHT OUTLET

(LIGHT GREY, FOR VACUUM UP TO 1 MBAR, SPECIAL 125 MM THREAD)

Ask for information

Ask for more information about:
CLASS II A2 BIOSAFETY CABINET – HIGH END Model

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.
Description
Class II biological safety cabinet in accordance with EN 12469. Meets the main requirements of NSF/ANSI 49, JIS K3800 and SFDA YY-0569. Protection of the user, the environment and the sample when working with biological agents at biosafety levels 1, 2 and 3

Ask for information

Ask for more information about:
CLASS II A2 BIOSAFETY CABINET – HIGH END Model

Before submitting the form, you can read the basic information on data protection here:

  • Responsible: TDI Diagnostic and Research Technology.
  • Purpose: Respond to your comment and notify you of new responses.
  • Legitimation: Your consent through this form.
  • Recipients: We do not plan to transfer your data to third parties, except by legal obligation.
  • Rights: You have the right to access, rectify, delete, portability and opposition to your data. The contact information will be stored on our servers.
  • Additional information: You can consult the additional and detailed information on data protection in the link at the bottom of this page.